ADAE-2 dental autoclave (Class B)

  • $1,500

[widgetic-facebook-like id=5a9bcc69ecb2a1d47b8b4569 autoscale=on width=431 height=62 resize=fill-width]


Specification:

Rated Voltage: 220V±22V,50Hz±1Hz(110V customized) 
vaporization temp:121℃/134℃
fuse: F15A ~250V
Ambient temp: +5℃~+40℃

 

Product Description:

1.LCD display
2.European Class B standard, 
3.With function of three pulse vacuum drying, 
4.Vacuum degree can reach to -0.82bar, 
5.Mechanical residual humidity less than 0.2%,
6.The jet steam generator with spiral pipe,
7.Door interlock protection, cavity with pressure, door cannot be opened make it more secure.
8.The built-in printer with USB is optional to record the disinfection process.
9.Water tank filled with warning system.

10.Fault defect alarm system

 

Chamber Size(mm):

18l:Φ250*355

23l:Φ250*465

 

Sterilization Program Process(three vacuum):

 

1:Press "Start" to enter the sterilization program;

 

2:Preheat: Program starts , the first detection of the pot wall temperature , such as the pot wall temperature is below 120 ℃ , the pot wall heating coil will be heated to a predetermined value 120 ℃.

 

3:Pulsating vacuum evaporator heating first evacuated to a vacuum pumped -82Kpa or when the pot holding a high temperature operation up time can not be pumped -82Kpa 4 minutes of rest ; when heated vacuum evaporator , heated to 170 ℃;

 

4:Steam injection Pump coupling solenoid valve is opened ; injection pump with some frequency to the water evaporator , evaporator water into steam injection sterilization chamber ( stainless steel pot ) .

 

5:Exhaust buck:

When the pressure inside the sterilization chamber reached 1OO Kpa, the exhaust valve opens, the pressure drops 10Kpa, the exhaust valve is closed.

 

6:Second pulsating vacuum:

Second evacuation, the degree of vacuum pumped -82Kpa sterilizing chamber maintained at high temperatures or when the state can not be pumped to run 4 minutes -82Kpa.

 

7:The second steam injection
Pump coupling solenoid valve is opened; injection pump with some frequency to the water evaporator, evaporator water into steam injection sterilization chamber (stainless steel pot).

 

8:The second exhaust buck

When the sterilization chamber pressure reaches 1OOKpa, the exhaust valve opens, the pressure drops 10Kpa, the exhaust valve is closed.

 

9:Third pulsating vacuum:

Second evacuation, the degree of vacuum pumped -82Kpa sterilizing chamber maintained at high temperatures or when the state can not be pumped to run 4 minutes -82Kpa time.

 

10.The third steam injection
Pump coupling solenoid valve is opened; injection pump with some frequency to the water evaporator, evaporator water into steam injection sterilization chamber (stainless steel pot).

 

11.Sterilization stage
Pressure, temperature control: 121 ℃ pressure control in 110Kpa, pressure and temperature for 20 minutes;
134 ℃ pressure control point 210Kpa, pressure and temperature for 4 minutes;

 

12.Buck exhaust exhaust valve opens, the pressure drops 10Kpa;

 

13:Drying stage (15 minutes) Open the vacuum pump, vacuum for 10 minutes after opening the inlet valve; the intake valve open off 1 minute; about 1 minute when the intake valve opens again; about 1 minute before closing; the last minute to stop all action;

 

14.Completion of sterilization.

 

 

---------Optional printer

 

Warranty 

One year