- Rated Voltage: 110V, 220V±22V, 50Hz±1Hz.
- vaporization temp:121℃/134℃.
- fuse: F15A ~250V.
- Ambient temp: +5℃~+40℃.
Watch the video for quick details
ADAE-5 autoclave digital touch (Class B)

Specification:
Product Description:
- Digital display.
- European Class B standard.
- With the function of three pulse vacuum drying.
- Vacuum degree can reach to -0.82bar.
- Mechanical residual humidity of less than 0.2%.
- The jet stream generator with a spiral pipe.
- Automatic door lock, procedure, and mechanical dual protection.
- Pressure power-off: Cavity with presure>5Kpa, Automatic door would power-off, and cannot be opened.
- The built-in printer with USB is optional to record the disinfection process.
- Water tank filled with a warning system.
- Fault defect alarm system.
Chamber Size(mm):
18L: Φ250*355.
23L: Φ250*465.
Warranty
one year
Optional part (Printer)

Sterilization Program Process(three vacuum):
1:Press "Start" to enter the sterilization program;
2:Preheat: The program starts, the first detection of the pot wall temperature, such as the pot wall temperature is below 120 ℃, the pot wall heating coil will be heated to a predetermined value 120 ℃.
3:Pulsating vacuum evaporator heating first evacuated to a vacuum pumped -82Kpa or when the pot holding a high-temperature operation up time can not be pumped -82Kpa 4 minutes of rest; when heated vacuum evaporator, heated to 170 ℃;
4:Steam injection Pump coupling solenoid valve is opened; an injection pump with some frequency to the water evaporator, evaporator water into steam injection sterilization chamber ( stainless steel pot).
5:Exhaust buck:
When the pressure inside the sterilization chamber reached 1OO Kpa, the exhaust valve opens, the pressure drops 10Kpa, the exhaust valve is closed.
6:Second pulsating vacuum:
Second evacuation, the degree of vacuum pumped -82Kpa sterilizing chamber maintained at high temperatures or when the state can not be pumped to run 4 minutes -82Kpa.
7:The second steam injection
Pump coupling solenoid valve is opened; an injection pump with some frequency to the water evaporator, evaporator water into steam injection sterilization chamber (stainless steel pot).
8:The second exhaust buck
When the sterilization chamber pressure reaches 1OOKpa, the exhaust valve opens, the pressure drops 10Kpa, the exhaust valve is closed.
9:Third pulsating vacuum:
The second evacuation, the degree of vacuum pumped -82Kpa sterilizing chamber maintained at high temperatures or when the state can not be pumped to run 4 minutes -82Kpa time.
10.The third steam injection
Pump coupling solenoid valve is opened; an injection pump with some frequency to the water evaporator, evaporator water into steam injection sterilization chamber (stainless steel pot).
11.Sterilization stage
Pressure, temperature control: 121 ℃ pressure control in 110Kpa, pressure, and temperature for 20 minutes;
134 ℃ pressure control point 210Kpa, pressure, and temperature for 4 minutes;
12. Buck exhaust valve opens, the pressure drops 10Kpa.
13:Drying stage (15 minutes) Open the vacuum pump, vacuum for 10 minutes after opening the inlet valve; the intake valve opens off 1 minute; about 1 minute when the intake valve opens again; about 1 minute before closing; the last minute to stop all actions.
14.Completion of sterilization.